Release 2026-08-27. Version 3.0.11.53, Revison M – With grey logo on cover
Functionality:
- ICUpilot and IscusCollect may be run on the ISCUSflex
- Built-in debug facilities for measurement curve assessment and more
- Enhanced printer functionality – adopted to A4 and Letter formats
- Possibility to print patient data to CSV files
- Possibility to select default Windows printer
- Operating system – Microsoft Windows 11
- Controlling computer – Surface Pro 11
New Functionality:
- Number of decimal places can be set manually for any analyte/catheter combination. Automatic scaling is still
available. - General dialog cleanup for all languages.
- Possibility to easily connect/disconnect a network share using the External media icon in the status bar.
- Corrected software defects (see end of Release notes)
Release 2026-03-25. Version 3.0.11.44, Revison M – With grey logo on cover
New functionality:
- ICUpilot and IscusCollect may be run on the ISCUSflex
- Built-in debug facilities for measurement curve assessment and more
- Enhanced printer functionality – adopted to A4 and Letter formats
- Possibility to print patient data to CSV files
- Possibility to select default Windows printer
- Operating system – Microsoft Windows 11
- Controlling computer – Surface Pro 11
Release 2026-09-14. Version 2.1.0.495 w patches, Revision L – With built-in receipt printer
Analyses approve more measurements (by adjusted parameters) than before:
- Glutamate BlankSlopeMin changed to -0.13 from -0.10 to handle differences in Glutamate batches
- DBUpdate can install/uninstall Glycerol Fix #3 (to handle Glycerol decreasing sensitivity over time)
Release 2022-03-14. Version 2.1.0.494, Revision L – With built-in receipt printer
Analyses approve more measurements (by adjusted parameters) than before:
- Eliminates Glucose blank and calibration problems
- Eliminates low concentration analyses issues
- An entry in the service log reveals values that are approved by the adjusted parameters
Control Samples LOT number gets stuck (2022-02-09)
In order to successfully cope with Control Samples nominal control levels a new Service Tool has been released. Especially for lot T27524, if Nominal Control Levels have been updated using a nominalcontrollevels.ini file, the lot number gets stuck and can’t be updated. To release the lot T27524 so i may be updated:
Download the file below, extract the zip-file to an empty USB stick and use it as a Service Tool and start UpdateDB according to the instructions below.
Lactate Calibration ResponseMax Fix (2021-11-11)
In order to cope with sensitive Lactate reagent batches a new Service Tool has been released. Download the file below, extract the zip-file to an empty USB stick and use it as a Service Tool to update the ResponseMax limit for Lactate calibrations.