Release 2026-08-27. Version 3.0.11.53, Revison M – With grey logo on cover

 

Functionality:

  • ICUpilot and IscusCollect may be run on the ISCUSflex
  • Built-in debug facilities for measurement curve assessment and more
  • Enhanced printer functionality – adopted to A4 and Letter formats
  • Possibility to print patient data to CSV files
  • Possibility to select default Windows printer
  • Operating system – Microsoft Windows 11
  • Controlling computer – Surface Pro 11

New Functionality:

  • Number of decimal places can be set manually for any analyte/catheter combination. Automatic scaling is still
    available.
  • General dialog cleanup for all languages.
  • Possibility to easily connect/disconnect a network share using the External media icon in the status bar.
  • Corrected software defects (see end of Release notes)

        Release 2026-03-25. Version 3.0.11.44, Revison M – With grey logo on cover

 

New functionality:

  • ICUpilot and IscusCollect may be run on the ISCUSflex
  • Built-in debug facilities for measurement curve assessment and more
  • Enhanced printer functionality – adopted to A4 and Letter formats
  • Possibility to print patient data to CSV files
  • Possibility to select default Windows printer
  • Operating system – Microsoft Windows 11
  • Controlling computer – Surface Pro 11

Release 2026-09-14. Version 2.1.0.495 w patches, Revision L – With built-in receipt printer

Analyses approve more measurements (by adjusted parameters) than before:

  • Glutamate BlankSlopeMin changed to -0.13 from -0.10 to handle differences in Glutamate batches
  • DBUpdate can install/uninstall Glycerol Fix #3 (to handle Glycerol decreasing sensitivity over time)
Unzip download file to an empty USB stick. Insert the same USB stick into the ISCUSflex and switch it on. Start the DBUpdate application and install Glycerol Fix #3 (GlycerolCalibRespMin = 1.2).

Release 2022-03-14. Version 2.1.0.494, Revision L – With built-in receipt printer

Analyses approve more measurements (by adjusted parameters) than before:

  • Eliminates Glucose blank and calibration problems
  • Eliminates low concentration analyses issues
  • An entry in the service log reveals values that are approved by the adjusted parameters
Instructions (updated 2022-03-28) Please finalize the update by following paragraph 3. Specific instructions for version 2.1.0.485 and later 

Control Samples LOT number gets stuck (2022-02-09)

In order to successfully cope with Control Samples nominal control levels a new Service Tool has been released. Especially for lot T27524, if Nominal Control Levels have been updated using a nominalcontrollevels.ini file, the lot number gets stuck and can’t be updated. To release the lot T27524 so i may be updated:
Download the file below, extract the zip-file to an empty USB stick and use it as a Service Tool and start UpdateDB according to the instructions below.

8050022P.zip

Instructions

Lactate Calibration ResponseMax Fix (2021-11-11)

In order to cope with sensitive Lactate reagent batches a new Service Tool has been released. Download the file below, extract the zip-file to an empty USB stick and use it as a Service Tool to update the ResponseMax limit for Lactate calibrations.

8050022O.zip

Instructions

 

Release notes for version 2.1.0.485 (released 2021-10-19)

  • More forgiving handling of Lactate, Pyruvate and Glycerol analyses
  • Increased linearity for Glycerol and Glutamate
  • Increased stability for Lactate
  • Pyruvate Level of Detection (LOD) increased to 10 µmol/L

Release notes for version 2.1.0.478 (released 2019-11-21)

  • Low sensitivity Pyruvate reagent handling improved

Release notes for version 2.1.0.477 (released 2019-09-20)

  • Dead lock situation when changing reagent cassette remedied
  • Rinsing Fluid LOT number and expiry date now recorded correctly in CLIA module

Release notes for version 2.1.0.475 (released 2019-08-27)

  • Urea calibration improvements
  • Issue when scanning Nominal Control Levels removed
  • Possibility for automatic save of service log

Release notes for version 2.1.0.472 (released 2019-04-11)

  • Exception when loading Nominal Control Levels removed

Release notes for version 2.1.0.471 (released 2018-09-20)

  • Possibility to LOAD Nominal Control Levels from the CLIA Control Levels dialog
  • A reminder about Nominal Control values not being consistent with entered auto control lot
  • Slovenian language added

Release notes for version 2.1.0.468 (released 2018-05-16)

  • Possibility to add a comment in the service log using a service code.
  • Possibility to set the low calibration linearity limit for Pyruvate to 300.

Release notes for version 2.1.0.465 (released 2017-10-12)

  • Improved handling of failing battery
  • Improved handling of high Pyruvate concentrations
  • Possibility for patient data archiving on other media
  • Improved handling of failing analyses

Release notes for version 2.1.0.459 (released 2016-09-13)

  • Possibility to send QC data to an external server (updated with Reagent and Calibrator LOTs)
    N.B. Requires update of IscusCollect to version 3.8.0.0 and ICUpilot to version 2.4.0.0
  • Cleaning cassette can be run in normal user mode
  • Possible to change patient ID
  • Possible to set patient data to final for easier handling
  • Return pump performance improved
  • Possible to wash the system from normal user mode
  • Service log and control samples print out now time and serial no stamped
  • Database dialog external media list box updates improved
  • Possibility to extend the service (PM) interval to 15 months

Updated 2022-08-23. Software for handling QC data from ISCUSflex (Download ISCUSflex QC Charts)
Installation Instructions (NOTE: Uninstall any previous version of the software)

  • View your QC data as Levey-Jennings charts
  • Get automated QC rules (according to Westgard) evaluation
  • View your QC data in ISCUSflex QC charts
  • Save you QC charts as image files
  • Tailor your QC data view as desired