Third generation Microdialysis Analyzer for point-of-care and advanced research

ISCUSflex is the third generation Microdialysis Analyzer for point-of-care and advanced research. This  analyzer uses enzymatic reagents and colorimetric measurements to monitor tissue chemistry from virtually any tissue or organ in the body.

The analyzer is easy to operate by medical professionals as well as researchers with a new and improved graphical user interface. Up to six different reagents provide unique opportunities for early detection of metabolic crisis and ischemia and to guide therapeutic interventions. The analysis results are displayed as trend curves for easy and fast interpretation. For comparison of other intensive care data the unique ICUpilot software is included.

ISCUSflex combines reliability and flexibility. It offers the opportunity to monitor up to eight patients at a time and is ideal for use in a clinical setting. With a capacity of 16 vials, the ISCUSflex offers batch analysis for advanced clinical and preclinical research. The throughput of the analyzer is approximately 30 measurements per hour.

Available for sale in Europe (CE-approval), US (FDA 510(k) clearance) and pending for launch in the rest of the world. For more info, please contact M Dialysis or your local distributor.

For the US version of ISCUSflex please click here

Microdialysis shall not be used as the sole basis for diagnosis or therapy.

  • Monitoring up to eight patients

  • Batch analysis capacity of 16 samples

  • Six available reagents: glucose, lactate, pyruvate, glycerol, glutamate and urea

  • Automatic calibration and quality control

PRODUCT NOTEPRODUCT NOTESIMPLE MAINTENANCESIMPLE MAINTENANCE

Technical information for ISCUSflex

Weight: 29 lbs/13 kg
Voltage: 100-240 V~, 50/60 Hz
Power consumption 100 VA
Dimensions 430 (H) x 350 (W) x 270 (D) mm
Type of protection: Class 1, Type B
Measurement principle Kinetic enzymatic analyzer
Vials: Microvials, Chromacol (CMA glass) and ETH (CMA plastic) vials
Samples: Microdialysates
Sample volume used: 0.4 – 2 µL/analysis (depending on analyte)
Min. sample volume: Sum of sample volumes per analyte + 2 µL
Reagent consumption: <15 µL/analysis (depending on analyte)
Pipetting imprecision: < 2 % (0.5 µL) rel. standard deviation
Calibration Automatic (every 6 hrs)
Warm-up time 10 minutes
Measuring time: 30 seconds
Throughput time/test: 90 seconds
Detector type Single beam filter photometer
Light source Class 1 LED
Wavelengths: 375 and 530 nm
Detector cell temperature: 99 ºF/37 ºC
Detector cell: Capillary flow cell 10 mm, 2 µL
Assay imprecision: <4 % rel. standard deviation for Control Sample Normal
Assay inaccuracy: <10 % for Control Sample Normal

Order information for ISCUSflex

NOTE: This product may not be approved by your national regulatory authorities.
Please contact your local representative for further information.
Ref. number
8003295 ISCUSflex Microdialysis Analyzer
8003296 ISCUSflex Microdialysis Analyzer, For Research Use Only